Angio ergometer with wheeled frame that makes it very easy to place the ergometer over a patient bed and move it. An rpm / watt meter is standard supplied with this ergometer.
Minimum load | 7 W | the minimum load the ergometer can provide | |
Maximum peak load | 1000 W | the maximum load the ergometer can provide for a short period of time | |
Minimum load increments | 1 W | the smallest steps with which the load can be added | |
Maximum continuous load | 750 W | the maximum power the ergometer can deliver continuous | |
Hyperbolic workload control | control of workload in hyperbolic way | ||
Linear workload control | control ergometer in a linear way | ||
Fixed torque workload control | control ergometer in a fixed torque way | ||
Maximum rpm independent constant load | 150 rpm | rpm as to which constant load can be applied | |
Minimum rpm independent constant load | 30 rpm | rpm as from which constant load can be applied | |
Optional heart rate controlled workload | control the device optionally through a set heartrate | ||
Electromagnetic "eddy current" braking system | system that very accurately applies a brake to the ergometer | ||
Dynamic calibration | continuous calibration during application of workload |
Workload accuracy from 7 to 100 W | 3 W | accuracy of the ergometer between 7 and 100 Watt | |
Workload accuracy from 100 to 500 W | 3 % | accuracy of the ergometer between 100 and 500 Watt | |
Workload accuracy from 500 to 1000 W | 5 % | accuracy of the ergometer between 500 and 1000 Watt |
Readout Distance | Readout the distance on control unit | ||
Readout RPM | Readout the actual rpm on control unit | ||
Readout Heartrate | Readout the actual heartarte on control unit | ||
Readout target HR | Readout target heartarte on control unit (only possible with HR option) | ||
Readout Energy | Readout the energy on control unit | ||
Readout Torque | Readout torque on control unit | ||
Readout Time | Readout time on control unit | ||
Readout Power | Readout the actual power on control unit | ||
Set Display | Choosing parameters to be shown on the display during the stress test | ||
Set Resistance | Possibility to define a specific resistance for each test subject | ||
Set P-Slope | Possibility to set the load regulation resistance | ||
Set Mode | Possibility to define parameters to be adjusted during the stress test | ||
Manual operation mode | Operate manually | ||
Preset protocol operation mode | Operate via preset protocols | ||
Analog operation mode | operate analog | ||
Terminal operation mode | operate with an external device | ||
External control unit | a control unit is connected with a cable to the ergometer | ||
Selfdesigned protocol operation mode | operate via selfdesigned protocols |
Analog connector | read-out of ergometer load in an analog way |
Maximum table width | 1090 mm | 42.9 inch | maximum distance between the outside of the instrument rails on a patient table |
Minimum table width | 760 mm | 29.9 inch | minimum distance between the outside of the instrument rails on a patient table |
115 V AC 50/60 Hz (138 VA) | operational power | ||
230 V AC 50/60 Hz (138 VA) | operational power |
IEC 60601-1:2005 | the product is IEC 60601-1 3rd edition compliant | ||
ISO 13485:2016 compliant | Lode fulfils ISO 13485:2016 requirements | ||
ISO 9001:2015 compliant | Lode fulfills ISO 9001: 2015 requirements |
CE class Im according to MDD93/42/EEC | CE certified | ||
CB according to IECEE CB | CB certified |