This option enables accurate and reliable automatic blood pressure measurement during cardiac rehabilitation, not disturbed by motion artifacts. The integrated BPM option is supplied as part of a complete cardiac rehabilitation setting using LCRM. This means that the BP measurement can be started and saved from the LCRM software. BP measurementis done according to the Korotkoff method
Note: for the blood pressure measurement, a trigger is needed
Accuracy BPM <300 mmHg | 3 mmHg | Accuracy of the BPM module between 0-300 mmHg | |
Minimum systolic measurement | 50 mmHg | Minimum BPM Measurement Range | |
Maximum systolic measurement | 250 mmHg | Maximum BPM Measurement Range | |
Minimum diastolic measurement | 20 mmHg | Minimum BPM Measurement Range | |
Maximum diastolic measurement | 150 mmHg | Maximum BPM Measurement Range |
TTL input | 3-5 V | TTL input required for BPM | |
Pulse length | 45 - 95 ms | Pulse length required for BPM | |
Maximum delay between RS-peak and pulse | 40 ms | Required for BPM | |
Variances in delay time (jitter) | <5 ms | Required for BPM |