The Gold Standard in Ergometry: with proven accuracy and reliability, the Excalibur sport is renowned worldwide as “the gold standard in ergometry”. The newly designed and improved Excalibur sport ergometer meets the latest requirements of modern sports medicine and research. Since athletes are becoming more and more powerful and testing more advanced than ever, this ergometer has been developed for extreme workloads up to 2500 watt! The new design ensures maximum stability at these high workloads. Thanks to the increased adjustability, versatile positioning of the test subject has never been better! The function of this product can be enhanced by using it in combination with our Lode Ergometry Manager software.
Minimum load | 8 W | the minimum load the ergometer can provide | |
Maximum peak load | 2500 W | the maximum load the ergometer can provide for a short period of time | |
Isokinetic workload control | control ergometer in isokinetic way | ||
Minimum load increments | 1 W | the smallest steps with which the load can be added | |
Maximum continuous load | 1500 W | the maximum power the ergometer can deliver continuous | |
Hyperbolic workload control | control of workload in hyperbolic way | ||
Linear workload control | control ergometer in a linear way | ||
Fixed torque workload control | control ergometer in a fixed torque way | ||
Maximum rpm independent constant load | 180 rpm | rpm as to which constant load can be applied | |
Minimum rpm independent constant load | 25 rpm | rpm as from which constant load can be applied | |
Optional heart rate controlled workload | control the device optionally through a set heartrate | ||
Electromagnetic "eddy current" braking system | system that very accurately applies a brake to the ergometer | ||
Dynamic calibration | continuous calibration during application of workload |
Workload accuracy below 100 W | 2 W | accuracy of the ergometer below 100 Watt load | |
Workload accuracy over 1500 W | 5 % | accuracy of the ergometer >1500 Watt | |
Workload accuracy from 100 to 1500 W | 2 % | accuracy of the ergometer between 100 and 1500 Watt |
Toeclips on pedals | fixation of the feet on pedals to enable optimal force transition | ||
Q-factor | 147 mm | Q-factor | |
Minimum leg length user | 725 mm | 28.5 inch | leg length to fit to pedals |
Minimum leg length user (incl. adjustable pedals) | 650 mm | 25.6 inch | leg length to fit to pedals when adjustable cranks are used (measured inseam) |
Vertical seat adjustment maximum | 938 mm | 36.9 inch | seat adjustment maximum height from crank axis to seat top |
Vertical seat adjustment minimum | 550 mm | 21.7 inch | seat adjustment minimum height from crank axis to seat top |
Horizontal seat adjustment minimum | 72 mm | 2.8 inch | horizontal seat adjustment minimum distance from crank axis to seat midpoint |
Horizontal seat adjustment maximum | 324 mm | 12.8 inch | horizontal seat adjustment maximum distance from crank axis to seat midpoint |
Seat angle adjustment maximum | 10 ° | angle adjustment of the seat bar versus the crankaxis | |
Allowed user weight | 180 kg | 396.8 lbs | maximum patient weight |
Horizontal handlebar adjustment minimum | 229 mm | 9 inch | horizontal handlebar adjustment minimum distance from crank axis to center of handlebar mounting point |
Horizontal handlebar adjustment maximum | 600 mm | 23.6 inch | horizontal handlebar adjustment maximum distance from crank axis to center of handlebar mounting point |
Vertical handlebar adjustment minimum | 465 mm | 18.3 inch | vertical handlebar adjustment minimum distance from crank axis to center of handlebar mounting point |
Vertical handlebar adjustment maximum | 855 mm | 33.7 inch | vertical handlebar adjustment maximum distance from crank axis to center of handlebar mounting point |
Handlebar adjustment angle | 360 ° | angle over which the handlebar can rotate |
Readout Distance | Readout the distance on control unit | ||
Readout RPM | Readout the actual rpm on control unit | ||
Readout Heartrate | Readout the actual heartarte on control unit | ||
Readout target HR | Readout target heartarte on control unit (only possible with HR option) | ||
Readout Energy | Readout the energy on control unit | ||
Readout Torque | Readout torque on control unit | ||
Readout Time | Readout time on control unit | ||
Readout Power | Readout the actual power on control unit | ||
Set Display | Choosing parameters to be shown on the display during the stress test | ||
Set Resistance | Possibility to define a specific resistance for each test subject | ||
Set P-Slope | Possibility to set the load regulation resistance | ||
Set Mode | Possibility to define parameters to be adjusted during the stress test | ||
Manual operation mode | Operate manually | ||
Preset protocol operation mode | Operate via preset protocols | ||
Analog operation mode | operate analog | ||
Terminal operation mode | operate with an external device | ||
External control unit | a control unit is connected with a cable to the ergometer | ||
Selfdesigned protocol operation mode | operate via selfdesigned protocols |
Analog connector | read-out of ergometer load in an analog way | ||
RS232 in connector | connection via RS232 | ||
RS232 out connector | connection via RS232 |
Product length (cm) | 130 cm | 51.2 inch | length of product in cm |
Product width (cm) | 70 cm | 27.6 inch | width of the product in cm |
Product height | 89 cm | 35 inch | height of product |
Product weight | 100 kg | 220.5 lbs | the weight of the product |
Power cord length | 250 cm | 98.4 inch | length of powercord including plugs |
Power cord IEC 60320 C13 with CEE 7/7 plug | European power cord and connector standard supplied with product | ||
Power cord NEMA | NEMA power cord and connector | ||
115 V AC 50/60 Hz (130 VA) | operational power | ||
230 V AC 50/60 Hz (130 VA) | operational power |
IEC 60601-1:2005 | the product is IEC 60601-1 3rd edition compliant | ||
ISO 13485:2016 compliant | Lode fulfils ISO 13485:2016 requirements | ||
ISO 9001:2015 compliant | Lode fulfills ISO 9001: 2015 requirements |
CE class Im according to MDD93/42/EEC | CE certified | ||
CE class of product with optional SpO2 | IIa | CE class of product with optional SpO2 | |
CE class of product with optional BPM | IIa | CE class of product with optional Blood Pressure Measurement | |
CB according to IECEE CB | CB certified |
Minimum operational temperature | 14 °C | minimum temperature at which the device will work within specification | |
Maximum operational temperature | 40 °C | maximum temperature at which the device will work within specification | |
Minimum operational air pressure | 70 kPa | minimum airpressure as from which the product will be operating within specification | |
Maximum operational air pressure | 106 kPa | maximum air pressure as from which the product will be operating within specification | |
Minimum operational non-condensing humidity | 30 % | minimum non-condensing humidity at which the product will operate within specification | |
Maximum operational non-condensing humidity | 90 % | maximum non-condensing humidity at which the product will operate within specification | |
Minimum storage & transport temperature | -25 °C | minimum temperature the product may be stored and transported at | |
Maximum storage & transport temperature | 70 °C | maximum temperature the product may be stored or transported at | |
Minimum air pressure storage & transport | 50 kPa | minimum air pressure the product may be stored or transported at | |
Maximum storage & transport air pressure | 106 kPa | maximum air pressure the product may be stored and transported at | |
Min. humidity storage & transport | 10 % | minimum non-condensing humidity that the product can be stored or transport at | |
Max. humidity storage & transport | 95 % | max. non-condensing humidity that the product can be stored or transported at |