Excalibur sport

Phased out - support up till 2029

The Gold Standard in Ergometry: with proven accuracy and reliability, the Excalibur sport is renowned worldwide as “the gold standard in ergometry”. The newly designed and improved Excalibur sport ergometer meets the latest requirements of modern sports medicine and research. Since athletes are becoming more and more powerful and testing more advanced than ever, this ergometer has been developed for extreme workloads up to 2500 watt! The new design ensures maximum stability at these high workloads. Thanks to the increased adjustability, versatile positioning of the test subject has never been better! The function of this product can be enhanced by using it in combination with our Lode Ergometry Manager software.

Specifications

Minimum load 8 W the minimum load the ergometer can provide
Maximum peak load 2500 W the maximum load the ergometer can provide for a short period of time
Isokinetic workload control
control ergometer in isokinetic way
Minimum load increments 1 W the smallest steps with which the load can be added
Maximum continuous load 1500 W the maximum power the ergometer can deliver continuous
Hyperbolic workload control
control of workload in hyperbolic way
Linear workload control
control ergometer in a linear way
Fixed torque workload control
control ergometer in a fixed torque way
Maximum rpm independent constant load 180 rpm rpm as to which constant load can be applied
Minimum rpm independent constant load 25 rpm rpm as from which constant load can be applied
Optional heart rate controlled workload
control the device optionally through a set heartrate
Electromagnetic "eddy current" braking system
system that very accurately applies a brake to the ergometer
Dynamic calibration
continuous calibration during application of workload
Workload accuracy below 100 W 2 W accuracy of the ergometer below 100 Watt load
Workload accuracy over 1500 W 5 % accuracy of the ergometer >1500 Watt
Workload accuracy from 100 to 1500 W 2 % accuracy of the ergometer between 100 and 1500 Watt
Toeclips on pedals
fixation of the feet on pedals to enable optimal force transition
Q-factor 147 mm Q-factor
Minimum leg length user 725 mm 28.5 inch leg length to fit to pedals
Minimum leg length user (incl. adjustable pedals) 650 mm 25.6 inch leg length to fit to pedals when adjustable cranks are used (measured inseam)
Vertical seat adjustment maximum 938 mm 36.9 inch seat adjustment maximum height from crank axis to seat top
Vertical seat adjustment minimum 550 mm 21.7 inch seat adjustment minimum height from crank axis to seat top
Horizontal seat adjustment minimum 72 mm 2.8 inch horizontal seat adjustment minimum distance from crank axis to seat midpoint
Horizontal seat adjustment maximum 324 mm 12.8 inch horizontal seat adjustment maximum distance from crank axis to seat midpoint
Seat angle adjustment maximum 10 ° angle adjustment of the seat bar versus the crankaxis
Allowed user weight 180 kg 396.8 lbs maximum patient weight
Horizontal handlebar adjustment minimum 229 mm 9 inch horizontal handlebar adjustment minimum distance from crank axis to center of handlebar mounting point
Horizontal handlebar adjustment maximum 600 mm 23.6 inch horizontal handlebar adjustment maximum distance from crank axis to center of handlebar mounting point
Vertical handlebar adjustment minimum 465 mm 18.3 inch vertical handlebar adjustment minimum distance from crank axis to center of handlebar mounting point
Vertical handlebar adjustment maximum 855 mm 33.7 inch vertical handlebar adjustment maximum distance from crank axis to center of handlebar mounting point
Handlebar adjustment angle 360 ° angle over which the handlebar can rotate
Readout Distance
Readout the distance on control unit
Readout RPM
Readout the actual rpm on control unit
Readout Heartrate
Readout the actual heartarte on control unit
Readout target HR
Readout target heartarte on control unit (only possible with HR option)
Readout Energy
Readout the energy on control unit
Readout Torque
Readout torque on control unit
Readout Time
Readout time on control unit
Readout Power
Readout the actual power on control unit
Set Display
Choosing parameters to be shown on the display during the stress test
Set Resistance
Possibility to define a specific resistance for each test subject
Set P-Slope
Possibility to set the load regulation resistance
Set Mode
Possibility to define parameters to be adjusted during the stress test
Manual operation mode
Operate manually
Preset protocol operation mode
Operate via preset protocols
Analog operation mode
operate analog
Terminal operation mode
operate with an external device
External control unit
a control unit is connected with a cable to the ergometer
Selfdesigned protocol operation mode
operate via selfdesigned protocols
Analog connector
read-out of ergometer load in an analog way
RS232 in connector
connection via RS232
RS232 out connector
connection via RS232
Product length (cm) 130 cm 51.2 inch length of product in cm
Product width (cm) 70 cm 27.6 inch width of the product in cm
Product height 89 cm 35 inch height of product
Product weight 100 kg 220.5 lbs the weight of the product
Power cord length 250 cm 98.4 inch length of powercord including plugs
Power cord IEC 60320 C13 with CEE 7/7 plug
European power cord and connector standard supplied with product
Power cord NEMA
NEMA power cord and connector
115 V AC 50/60 Hz (130 VA)
operational power
230 V AC 50/60 Hz (130 VA)
operational power
IEC 60601-1:2005
the product is IEC 60601-1 3rd edition compliant
ISO 13485:2016 compliant
Lode fulfils ISO 13485:2016 requirements
ISO 9001:2015 compliant
Lode fulfills ISO 9001: 2015 requirements
CE class Im according to MDD93/42/EEC
CE certified
CE class of product with optional SpO2 IIa CE class of product with optional SpO2
CE class of product with optional BPM IIa CE class of product with optional Blood Pressure Measurement
CB according to IECEE CB
CB certified
Minimum operational temperature 14 °C minimum temperature at which the device will work within specification
Maximum operational temperature 40 °C maximum temperature at which the device will work within specification
Minimum operational air pressure 70 kPa minimum airpressure as from which the product will be operating within specification
Maximum operational air pressure 106 kPa maximum air pressure as from which the product will be operating within specification
Minimum operational non-condensing humidity 30 % minimum non-condensing humidity at which the product will operate within specification
Maximum operational non-condensing humidity 90 % maximum non-condensing humidity at which the product will operate within specification
Minimum storage & transport temperature -25 °C minimum temperature the product may be stored and transported at
Maximum storage & transport temperature 70 °C maximum temperature the product may be stored or transported at
Minimum air pressure storage & transport 50 kPa minimum air pressure the product may be stored or transported at
Maximum storage & transport air pressure 106 kPa maximum air pressure the product may be stored and transported at
Min. humidity storage & transport 10 % minimum non-condensing humidity that the product can be stored or transport at
Max. humidity storage & transport 95 % max. non-condensing humidity that the product can be stored or transported at
*Specifications are subject to change without notice.

Science

FAQ

The Corival handlebar is adjustable within 0-360 degrees.
The Excalibur Sport's handlebar is adjustable in both vertical (465- 855 mm) and horizontal (229- 600 mm) direction.
Yes, you can use most commercial available saddles. For the Excalibur an easy saddle exchange option is available. Please contact your service coordinator
Yes, you can find interface sheets for most common devices on the website at the "interfacing" tab of the product. If your CPET device is not present you can ask us to support you. In most cases we can quickly provide you with a solution.
Yes, this is possible. Contact your local supplier for more information.
The Torque factor is used to calculate braking Torque during Wingate sprinttest. Torque factor x body weight = braking Torque this is used in this test. Lode is using 0.7 as the default Torque factor for normal healthy test subjects. Athletes will need a higher Torque factor to get their optimal braking torque, women slightly lower, testsubjects with severe obesity, muscle atrophy, muscle dystrophy etc will need a lower Torque factor (see lt. Bar Or).
Your Lode ergometer is standard supplied with one display. The contrast of this display is factory set and can adjusted by qualified service personal only. If your ergometer is supplied with a programmable option the contrast of this display can be set as follows: go to SYSTEM PARAMETERS select SETTINGS go to CONTRAST. Change the visability according your wishes and save the changes.
Yes you can request the actual pedalling speed and power when you are cycling in Manual, Analogue or Terminal mode.
The load will only be returned when the pedalling speed is 30 RPM or more. For technical details see the Lode RS232 protocol
The Q factor of a bicycle ergometer is the distance from crank to crank. For the Excalibur the Q factor is simular to a racing bike meaning 147mm (+/- 3 mm).
The Q factor of the other ergometers is a little bit bigger (like a city bike), for the Corival this is 180 mm.
Yes you can. Make sure the chipset used in the converter is from FTDI to have good data acquisition.
Not sure what this means? Contact us and we will help you to get the right converter.
SPD one sided, It is possible to use one side for normal (sport)shoes and the 'click' side for Shimano shoes.
Yes, Due to the fact that the scientific research on the spread of Covid-19 is not yet conclusive Lode recommends to disinfect and clean all Lode products and their accessories (cuffs, heart rate belts, SpO2 sensors, handgrips/rails, touchscreens etc) that are in contact with the patient and or the medical personnel with a standard disinfectant as used in the hospital before and after every patient, with attention and care!

Since these agents are sometimes aggressive it might be advisable to check the effect on sensitive surfaces, like cushioning on non-visible areas. Lode will not take responsibility for discoloration due to these agents.

Part number

  • 925900

Certificates

ISO 13485: 2016 Certificate
CE Certificate Stress Test and Force Measurement Devices
ISO 9001:2015 Certificate
IEC 60601-1:2005 – ed. 3 + C1 + C2 (cycle ergometers) certificate
FDA Registration certificate
Manufacturers Declaration of Conformity 27 Lode
UL CSA Corival series certificate

LEM Demo Software

955900 LEM 10.15.0 Demo software

Manuals

Manual Excalibur Sport German
Manual Excalibur Sport English
Manual Excalibur Sport French
m925900 manual excalibur sport hungarian
Manual Excalibur Sport Italian
Manual Excalibur Sport Dutch
Manual Excalibur Sport Norwegian
Manual Excalibur Sport Russian
Manual Excalibur Sport Swedish

User guide

User guide Excalibur Sport English

Brochures

Brochure Excalibur sport English
pdf
Brochure Cardio Pulmonary English
pdf
Brochure Excalibur sport German
pdf
Brochure Cardio Pulmonary German
pdf
Brochure Excalibur sport Spanish
pdf
Brochure Cardio Pulmonary Spanish
pdf

Posters

Poster Excalibur sport English
Poster Cardio Pulmonary English
Poster Excalibur sport German
Poster Cardio Pulmonary German
Poster Excalibur sport Spanish
Poster Cardio Pulmonary Spanish

Service

Part number

  • 925900

Interfacing protocols

AEI Technologies
AMEDTEC Medizintechnik Aue GmbH
BTL
CareFusion
Cognimed
Cortex
Custo Med
Delmar Reynolds
Edan Instruments Inc.
Enverdis GmbH
ET Medical
GE Healthcare
Korr Medical Technologies Inc
Labtech
Lode
Med-tronik
Medical Graphics
Mesa Medizintechnik
Mortara
Nihon Kohden
Norav
nSpire Health
Parvomedics
Philips Medical
Progetti
Pulse Biomedical, Inc.
Quinton
Schiller
Seiva
Welch Allyn
1350K
1500 K / 1550K - Cardiofax V
600 USB system
9320K
AT-10
AT-104 PC
AT-110
AT10 Plus
BreezeSuite Software
BTL-08 ECG
Cardiax PC-EKG
Cardio Coach Monitor
Cardio Collect 12
Cardiologic Explorer 12C Ergo
CardioPerfect
Cardiosoft
Cardiosoft + Corina
Cardiosoft MAC 1200 ST
CS200
Cubestress
Custo Diagnostic Stress ECG
EC12 R/S
ECG Pro
Edan EC-12 EXPRESS
Edan ST-1212 STRESS ECG
Ergolog
GE Case
GE Case 8000
GE Case Basic
GE Case Performance
GE Case Premium
GE Case Value
Lactware Software
Lode Ergometry Manager
MAC 5000
MAC 5500st
MasterScreen® ECG / Viasys Oxycon Pro
Metasoft
MOXUS Modular Metabolic System
Multi Control
MyStress system
PC-ECG 1200
Q-Stress
QRS-card
StressVue
TM Cardio 1000/2000
TrueOne
X-scribe
XR-200